Search
Key Features
  • MCERTS Approved
  • Single analyzer capable of measuring five components
  • Compact and lightweight
  • Built-in sampling unit
  • Simple to operate, easy to read
  • Optional electronic cooler

This latest gas analyzer is both portable and MCERTS approved. Capable of measuring five components, the new Horiba PG-250 is a highly reliable and versatile gas analyzer for compliance testing of NOx, SO2, CO, CO2, and O2. Housed in a single lightweight and fully portable case, it utilizes the same measurement principles as permanently installed CEMS - unlike other portable gas analyzers that rely upon electro-chemical sensors.

The Horiba PG-250 rents with stainless steel sample probe, ds-200 drain seperator, power cord, sample gas filter element for probe filter, tip filter for reference gas, output signal cable Nox, SO2, CO2, CO and O2, plastic dust cover for PG 250, tygon tubing and instruction manual.

Applications
  • Emissions testing
  • Combustion Efficiency testing
  • Relative accuracy test audits
  • Pollution control systems
  • Power generation & refineries
  • Gas turbines & boilers
  • Incinerators
  • Electric utilities
Technical Specification
Components Measured: Nitrogen Oxide (NOx), Sulphur Dioxide (SO2), Carbon Dioxide (CO2), Carbon Monoxide (CO), and Oxygen (O2)
Principle of Measurement Chemiluminescence measurement for Nox, Non-Dispersive Infra Red Absorption (SO2, CO2, CO)
Galvanic cell measurement for O2
Repeatability ±0.5% F.S. (Nox:>100 ppm; CO:>1000 ppm
±1.0 F.S. (other components and ranges)
Linearity ±2.0% F.S.
Drift ±1.0% F.S. per day
±2.0% F.S. per day for SO2
Response Time 45 seconds (moving average from inlet is 10 sec)
Only SO2: 240 seconds with selective moving average of 10 sec or 30 sec.
Sample Gas Flow Rate Approximately 0.4 SLPM
Ambien Humidity <85% R.H.
Ranges: NOx: 0-25/50/100/250/500/1000/2500ppm
SO2: 0-200/500/1000/3000 ppm
CO: 0-200/500/1000/2000/5000ppm
CO2: 0-5/10/20 vol%
O2: 0-5/10/25 vol%
Display Measured value (3 or 4 digits) active, range, flow rate
Ambient Temperature 5-40°C; 41-104°F
Power Requirements 100-120 V ac; 200-240 V ac at 50/60 Hz
Power Consumption 250 VA/400 VA
Sample Gas Conditions Temperature: ,<40°C (<104°F) at Analyzer Inlet: H2O:Dust:<0.1 g/Nm3, Pressure: ±0.98 kPa (±100mm H2O)
Output: DC 4-20mA (Nonisolated), RS-232C Interface
Dimensions
Title (mm) (inch) (kg) (lbs)
260 x 260 x 510 mm 10.2 x 10.2" x 20.0"" 17Kg 37.5Lbs